Assessment of metal artifacts from titanium wrist prostheses: photon-counting versus energy-integrating detector CT

Scritto il 01/05/2025
da Nina Kämmerling

Eur Radiol Exp. 2025 May 1;9(1):45. doi: 10.1186/s41747-025-00587-w.

ABSTRACT

BACKGROUND: We compared photon-counting detector computed tomography (PCD-CT) polyenergetic images, PCD-CT virtual monoenergetic images (VMI), and energy-integrating detector computed tomography (EID-CT) polyenergetic images regarding bone visualization and metal artifacts in patients with titanium wrist prostheses.

METHODS: After ethical approval, 15 patients were examined with PCD-CT and EID-CT. Polyenergetic images were reconstructed, as well as 130-keV VMI for PCD-CT. Five radiologists evaluated bone visualization, interpretability at metal-bone interface and metal artifacts using a 7-point ordinal scale. Streak artifacts and artifacts at the bone-metal interface were quantitatively assessed. Differences between image setups were analyzed using Friedman test and one-way ANOVA with post hoc tests.

RESULTS: Bone visualization was superior in PCD-CT polyenergetic images (median rating 6, range 3-7) compared with VMI (5, 3-7; p < 0.001) and EID-CT (5, 3-7; p = 0.018). Streak artifacts were more pronounced with PCD-CT polyenergetic images (4, 3-6) compared with EID-CT (5, 4-6; p = 0.003) and PCD-CT VMI (5, 3-7; p = 0.002), with quantitative results showing least streak artifacts in PCD-CT VMI, followed by EID-CT and PCD-CT polyenergetic images (50 ± 7%, 70 ± 6%, and 79 ± 5%, respectively; p < 0.001). Interpretability at bone-metal interface was better with PCD-CT polyenergetic images (5, 2-7; p = 0.045) and EID-CT (5, 3-6; p = 0.018) compared with PCD-CT VMI (4, 2-6), without quantitative differences.

CONCLUSION: Streak artifacts from titanium wrist prostheses were reduced using 130-keV PCD-CT VMI, while bone visualization was highest using PCD-CT polyenergetic images.

RELEVANCE STATEMENT: In patients with wrist implants, photon-counting detector CT allows for effective metal artifact reduction using virtual monoenergetic images and improved bone visualization using polyenergetic images. As polyenergetic images and VMI have different advantages, access to both image setups may benefit diagnostic evaluation.

KEY POINTS: Virtual monoenergetic images (VMI) presented a substantial reduction of metal streak artifacts. Polyenergetic images exhibited better image quality for bone imaging compared with VMI. A combination of image reconstructions should be preferred depending on the diagnostic task.

PMID:40310571 | DOI:10.1186/s41747-025-00587-w